4. Installing the equipment
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
14
No.600-2111-09
Put HT-300A on the top plate.
Insert and pull out the tube from the hole.
If the tube cannot pull out smoothly, replace HT-
300A.
Connect the Teflon tube to the waste-vial through
the spacer.
Pull the tube from the Pump side and house the
vial in the holder.
* Put the vials on right and left side, and check the
height. When the bottle has floated, please roll and
put in a little seal-ring tapes.
Connect the tube to the Pump.
* The waste bottle is not attached.
Connect two signal cables and power cable, from the PE-1.
(5) Connect the two signal cables and the
power cord of the PE-1 to the HT-
300A. And turn on the switch.
Waste
vial
Spacer
FLW-30-6RU3-C
PTFE 2X3
PTFE 2x4
Drain tunk
(Not attached)
Switch