6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
56
No.600-2111-09
Adjustment of the sample position
If the sample changer does not work properly even when the power is turned on,
please do the position adjustment.
(1) Turn on the power while pressing the SET key to display the following screen.
(2)
Press the MET and ENT key at the same time.
(3)
Select “BORDERS MARK-OUT”
(4) Remove the syringe from the HT-300A, and set the waste bottle, and the vials to
No. 48, 50 of the turntable. And then press the START key.
Borders mark-out will begin.
(5) Adjust positon of the turret, so that the syringe guide is above the waste bottle,
using
key. And then down the needle
using key.
Display of [STO] appears when
lowered to below the needle. Press
ENTER key.