7. PE1000Win software
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
81
No.600-2111-09
Measurement condition (Control)
The standard check function and blank check function can be used to monitor the
equipment during unattended operation.
To set these functions, select [Run]-[Mode] from the menu and click the "Control" tab.
The standard check function can check the equipment status by periodically measuring a
standard sample when an unknown sample is measured.
The blank check function allows measurement after a high concentration is measured and
then the blank is reduced.
Standard check
To execute the standard check, preset the "STD" conditions in the following window in
advance.
STD [pg]
Enter the standard value to be measured in the standard check.
Max Dev [%]
Specify a deviation (%) from the standard value. This is used for
judging whether a measured standard sample is abnormal or not.
Abnormal Condition Specify whether to continue or cancel the measurement when an
abnormality is detected.