7. PE1000Win software
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
77
No.600-2111-09
Pre-Wash Solvent
Specify the place of the liquid that washes the syringe before dispensing a sample.
Pre wash count
Specify the number of times to wash the syringe before dispensing a sample.
When “0” is specified, it will not be washed.
Pre wash volume (uL)
Specify the amount of the liquid to be used to wash the syringe before dispensing a
sample. An amount equal to or greater than the syringe size cannot be specified.
Post-Wash1 Solvent
Specify the place of the liquid that washes the syringe after injecting a sample.
Post wash1 count
Specify the number of times to wash the syringe after injecting a sample.
When “0” is specified, it will not be washed.
Post wash1 volume (uL)
Specify the amount of liquid to be used to wash the syringe after injecting a sample.
An amount equal to or greater than the syringe size cannot be specified.
Post-Wash2 Solvent
Specify the place of the liquid that secondarily washes the syringe after injecting a
sample.
Post wash2 count
Specify the number of times to wash secondarily the syringe after injecting a sample.
When “0” is specified, it will not be washed.
Post wash2 volume (uL)
Specify the amount of liquid to be used to wash secondarily the syringe after injecting a
sample.
An amount equal to or greater than the syringe size cannot be specified.
Fill volume (uL)
Specify the amount of a sample with which the syringe is washed before taking the
sample.
Fill speed (uL/s)
Specify the speed at which a sample is sucked into the syringe.
For low boiling point or high viscosity samples specify a sufficiently low speed.
Sample wash Count
Specify the number of times to wash the syringe with a sample before taking the sample.
When “0” is specified, it will not be washed with the sample.
Pullup strokes
Specify the number of times to perform pull-ups in order to degas the syringe.
The sample is also agitated in the vial at the same time.
When “0” is specified, no pull-ups will be performed.