6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
42
No.600-2111-09
Replacing the syringe
(1) Press the EXIT key of the HT-300A to display the following title screen.
(2) Press the SET key while the title screen is being displayed.
(3) Press the
key to move to [Vial/Syringe Setup]. And then press the
key for
going into the [Vial/Syringe Setup] menu.
(4) Press the
key to move to [Syringe Install]. And then press the
key for going
into the [Syringe Install] menu.