6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
55
No.600-2111-09
Manual operation of the syringe
Use this when manually moving the position of the syringe to replace the syringe, etc.
(1) Press the EXIT key of the HT-300A to display the following title screen.
(2) Press the SET key while the title screen is being displayed.
(3) Press the
↑
key to move to [Manual Operations], and press the
→
key to select [Manual
Operations].
(4) Use
←
and
→
keys to select the relevant item, and use the
↑
and
↓
keys to manually
move it.
PLG: Vertical movement of the plunger
NDL: Vertical movement of the syringe
TUR: Horizontal swinging movement of the tower
TRY: Rotating movement of the sample tray
(5) After operation, press the EXIT key to exit the manual operation mode.
(6) Press the SET key to return to the title screen.