2. Principles
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
5
No.600-2111-09
Table 1: Description of each part of the PE-1
No.
Name
Function
Remarks
1
Sample inlet
Sample inlet port
A septum is built in.
2
Sample heating
furnace H1
Catalytic furnaces
H2 and H3
Sample and catalytic furnaces
3
Jacket heater H4
Keeps the joint warm.
4
Flow meter
Indicates the carrier gas flow rate
(only used when checking for leaks).
0.5 L/min
5
Auxiliary power
switch
Includes the function of automatically
turning the heater off if it has
overheated.
6
Mercury collector
H5
Collects mercury in combustion gas.
7
Fluorescence cell
Fluorescence cell made of quartz
8
Mercury lamp
Mercury lamp
9
Phototube
For measuring the amount of the lamp
intensity
10
Gas-liquid
separation bottle
Separates any liquid from the
combustion gas.
11
Ar pressure
regulator
Ar pressure regulator
0.2 MPa
(Max. 1.0 MPa)
12
Air pressure
regulator
Air pressure regulator
0.2 MPa
(Max. 1.0 MPa)
13 Exhaust port
Outlet port for discharging exhaust gas
after measurement
Be sure to discharge
it outdoors.
14 PC terminal
RS232C terminal for connecting to PC
15 RH-PE terminal
Terminal for connecting the collection-
tube heating furnace
Optional
16
Power connection
terminal
Voltage can be selected from 100, 110,
120, 220, 230, or 240 V before
shipment.
1 KVA
17 Main power switch Includes a built in circuit breaker.
18 Mercury trap
Activated carbon for removing mercury
from Ar and the air