7.14.3
Selecting the load cell type
7.14.4
Calibration procedure
During calibration, no data is changed in the digital load cells. The calibration data and
parameters are saved in the instrument. The unique serial numbers of the connected load
cells are monitored.
For the calibration the following order must be followed:
-
Search for load cells, see Chapter
-
Assign load cells, see Chapter
.
4. Select [Baudrate] and confirm.
A selection window opens.
5.
Select "19200 bd" and confirm.
6. Select [Stopbits] and confirm.
A selection window opens.
7.
Select "1" and confirm.
8. Press
to exit the menu and to save the settings.
1.
Select
-[Weighing point]- [Weighing point A].
A selection window opens.
2.
Select "Pendeo Load Cells" and confirm.
3.
Press
to exit the menu and save.
Transmitter in field housing PR 5230
7 Commissioning
Minebea Intec
EN-164