Saving configuration and calibration data
Loading configuration and calibration data into the device
NOTICE
All data which can be configured in the Setup menu are overwritten!
If the file is loaded into several devices, changing the network settings and the host
name is indispensable!
9.1.4
Resetting the device to the factory settings
Note:
Reset to the factory settings is possible only, when the CAL switch is open. The IP
address and the Hostname remain unaffected.
1.
Click on [Backup] to create a backup .g. on the notebook.
2.
Click on [Save as].
3.
Create and open the required directory e.g. on the notebook.
4. Click on [Save] to save the backup file in the relevant directory.
1.
Open the CAL switch; see Chapter
.
2.
Click on [Select File] (depending on the Internet browser).
3.
e.g. on the notebook, navigate to the folder where the backup file was saved.
4. Select Backup file.
5.
Click on [Restore].
The selected file is loaded into the device.
6. Click on the
symbol in the Internet browser to return to the WEB menu.
7.
Close the CAL switch.
1.
Click on
.
2.
Click on
.
The following prompt window appears:
9 Extended functions
Transmitter in field housing PR 5230
EN-245
Minebea Intec