Serial number and current weigh of connected
load cells are displayed.
— New
Query window for new calibration: Corner
correction will be reset Yes, No
— Local gravity
Enter the on-site gravitational acceleration
(Standard: Hamburg 9.81379 m/s
2
)
— Number of
platforms
Only for Pendeo Truck: Only shown if the
number of load cells is equal to 8.
— Number of
vessel feet
Only for Pendeo process: Enter the number.
— Max
Enter maximum load:
0.000010
<3000>
9999998 <kg>, t, lb, g, mg,
oz
— Scale interval
Enter verification interval (1 d): <1>, 2, 5, 10, 20,
50 displayed according to the decimal places at
Max and the weight unit.
— Dead load
dead load: Weight of the empty scale
— CAL weight
Enter CAL weight: 0.000010
9999998 <kg>, t,
lb, g, mg, oz
— Corner
correction
Platform 1, platform 2 (only shown if the
number of load cells is = 8).
O.k., when realized.
— Modify
May be used only for minor changes (e.g.
changing the dead load). Otherwise, always use
[New]!
— Param
Parameter settings
— Ambient
conditions
Select environmental conditions: Very stable,
stable, unstable, very unstable
— W&M
Selection legal-for-trade mode: <none>, OIML,
NSC, NTEP
— Unbal.
Check deviat.
Average Deviation: The plausibility check is
activated when the average deviation is >0%.
Input: 0
100 %
— Standstill time
Input of the standstill period: 0.01 s
<0.50
s>
2.0 s (The range depends on the measuring
time.)
— Standstill
range
Input of the standstill range: 0.00 d
<1.00
d>
10.00 d (The range depends on the
measuring time.)
— Tare timeout
Input of timeout when there is no standstill: 0.1
s
<2.5 s>
25 s
— Zeroset range
Input of timeout period when there is no
standstill: 0.00 d
c
<50.00 d
c
>
10000.00 d
c
— Zerotrack
range
Input of the zerotrack range: 0.00 d
c
<0.25
d
c
>
10000.00 d
c
— Zerotrack step
Input of the zerotrack step: 0.00 d
c
<0.25
d
c
>
10.00 d
c
— Zerotrack time
Input zerotrack time: <0.0 s>
25 s
— Overload
Input of the weighing range above the
maximum load (Max) without error message:
0
<9 d
c
>
999999 d
c
7 Commissioning
Transmitter in field housing PR 5230
EN-121