By pressing [by mV/V] the value can be entered directly.
By pressing [Change] the selected linearization point can be changed.
By pressing [Delete] the selected linearization point can be deleted.
The window shows the set linearization points.
6. Select a linearization point, place the corresponding weight on the scale, and press
the [by load] softkey.
The value corresponding to the weight is automatically entered in mV/V.
7.
Repeat these steps to automatically enter the corresponding values for the weights
of all set linearization points in mV/V.
8. Press
to switch to the previous window.
Transmitter in field housing PR 5230
7 Commissioning
Minebea Intec
EN-138