Normally, calculation of the dead load (scale without load/empty hopper) is not
necessary.
Subsequent dead load correction (as described in Chapter
re-determination of the dead load, when the scale/hopper is empty.
Example:
-
1 load cell with nominal characteristic value C
n
= 2 mV/V
-
at maximum capacity of 2000 kg
-
Maximum load 1000 kg
-
Dead load 500 kg
-
Load cell supply voltage 12 V DC
7.12.9
Calibrating with load cell data (smart calibration)
If the scale is not used in legal metrology, calibration without weights can be performed.
The easiest method is the one using load cell data without calculation.
1.
Press the [by data] softkey.
Transmitter in field housing PR 5230
7 Commissioning
Minebea Intec
EN-134