Parameter table
Menu item
Selection
Description
[no output]
Analog output is unused.
[Gross D08]
Output of the gross weight.
[Net if tared D09]
Output of the net weight, if tared; otherwise gross
weight
[Selected D11]
Output of the gross/net value on the display, de-
pendent on SPM bit X72.
[Analog mode]
[Transparent D30]
Output of the value in D30
[0
20 mA]
Output of 0
20 mA.
[Analog range]
[4
20 mA]
Output of 4
20 mA.
[0 mA]
Set output to 0 mA.
[4 mA]
Set output to 4 mA.
[20 mA]
Set output to 20 mA.
[Output on error]
[hold]
The last output value is held.
[0 mA]
Set output to 0 mA.
[4 mA]
Set output to 4 mA.
[20 mA]
Set output to 20 mA.
[Output if < 0]
[linear]
The output drops below 4 mA down to the limit (at
4
20 mA).
[0 mA]
Set output to 0 mA.
[4 mA]
Set output to 4 mA.
[20 mA]
Set output to 20 mA.
[Output if > Max]
[linear]
Output increases above 20 mA until the limit is re-
ached.
[Weight at 0/4 mA]
Weight value for 0/4 mA output.
[Weight at 20 mA]
Weight value for 20 mA output.
Quit the menu and save parameters by pressing
.
Note:
Adapting the analog output, see Chapter
7.16 System information