If not all data was determined when calibrating with [New] (e.g. dead load not set/
entered), this message is shown:
Press [Yes] to confirm and then press again the
softkey, another prompt is displayed:
If you press [Undo], changes are not saved and the display returns to the selection menu
for the weighing points.
Leaving the menu is displayed by "Exit calibration".
7.12.15
Parameter Input
The menu is accessible via
-[Weighing point]- [Calib]- [Param].
7 Commissioning
Transmitter in field housing PR 5230
EN-141