— Zeroset range
± Range of zero point when there is no
standstill. Input: 0.00 d
<50.00 d>
10000.00
d
— Zerotrack range
Input of the zerotrack range: 0.00 d
<0.25
d>
10000.00 d
— Zerotrack step
Input of the zerotrack step: 0.00 d
<0.25
d>
10.00 d
— Zerotrack time
Input zerotrack time: <0.0 s>
25 s
— Overload
Input of the weighing range above the
maximum load (Max) without error message:
0
999999 d
— Minimum weight
Input of the minimum load: 0 d
<50
d>
999999 d
— Range mode
Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.12.15.2 and 7.12.15.3.
— Range limit 1
Input of the limit 1: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— Range limit 2
Input of the limit 2: In weight, unit same as Max,
transition from small to medium scale interval
Only for [Multiple range] or [Multi-interval]!
— View
(when CAL-switch is closed)
— Max
Display only
— Scale interval
Display only
— Dead load at
Display only
— Max at
Display only
— Calibrated at
Display only
— Sensitivity (µV/d)
Display only
— Param
Items as for [Param] (display only)
— Setup
Setup of the xBPI scale
— Calibration
Calibration of xBPI scale
— Dead load
Dead load/preload
— Set
Set dead load/preload: Accept, ResError, Abort
— Delete
Clear dead load/preload: Accept, ResError,
Abort
— Span
— Calibrate with user
weight
— Calibrate with auto
weight
— Adjust with default
weight
Calibrate with standard weight.
— Adjust with internal
weight
Calibrate with internal weight.
— Linearity
Setting linearity.
— Default
Resetting the Device to the Factory Settings:
Accept, ResError, Abort
— User
User defined settings: Accept, ResError, Abort
7.11.7.2
"xBPI scale" weighing point
Calibrate with user-defined weight.
Calibrate with automatic weight detection.
Transmitter in field housing PR 5230
7 Commissioning
Minebea Intec
EN-118