Specifications
Description
Data
Transmission rate
10 Mbit/s and 100 Mbit/s
Auto detection (100, FullDX)
Protocol
ProfiNet I/O
Connection mode
Network
Configuration
XML file
"GSDML-Vx.xx-Sartorius-PR5230-xxxxxx.xml"
Potential isolation
yes
Cable type
Twisted pairs, screened, e.g. patch cable CAT5 Autolink (stra-
ight or crossover)
Cable impedance
150
Ω
Cable length to HUB
Max. 115 m
Certificate
ProfiBus Nutzerorganisation e.V. for HMS Industrial Net-
works AB
Certificate No: Z10931
Note:
The IP address and the subnet mask are set at
- [Fieldbus parameter] (see also
Chapters
and
).
The XML file is stored on the CD (Fieldbus directory of the corresponding instrument)
supplied with the unit. The current file is also available for download via the Internet.
http://www.minebea-intec.com
Transmitter in field housing PR 5230
4 Device installation
Minebea Intec
EN-96