E:Sense
Error message: if no load cells are connected, see also
Chapter
E:NoCom
Error message: if there is no communication with the xB-
PI scale (see also Chapter
).
E:HardE
Error message: unable to read weight values from the
ADC (analog-digital converter); see also Chapter
E:NotRd
Error message: no load cells or no scale is connected (see
also Chapter
).
The weight display appears.
Set the date and time after turning on the device for the first time; see Chapter
7.3
Switching off the device
The device is switched off/disconnected from power supply by pulling the plug.
7.4
Warm-up time
The device requires a warm-up time of 30 minutes before calibration.
7.5
Finding and connecting a device with a notebook/PC
If the device is connected to a notebook/PC via a point-to-point connection, an IP
address is negotiated via function "AutoIP". This can take up to 2 minutes!
NOTICE
When the IT/DHCP network cable is temporarily connected between the notebook/
PC and a device, the DHCP server is lost and the notebook/PC returns to the auto-IP
address within approx. two minutes!
Reason: The DHCP server/client relationship is checked cyclically at 2
3-minute
intervals.
1.
On the notebook/PC, set the LAN local and Internet Protocol properties to "Obtain an
IP address automatically" depending on the operating system.
2.
On the device, under
- [Network parameter], activate the "Use DHCP" parameter
(default/factory setting).
7 Commissioning
Transmitter in field housing PR 5230
EN-109
Minebea Intec