NOTICE
The
rotary switch settings will not be used.
Ensure that the three rotary switches for node address 1
99 are set to posi-
tion "0."
Settings are defined via
- [Fieldbus parameter][Profibus-DP].
4.6.5.2
Status indicator
Requirements:
-
The items are defined via
- [Display items]- [Fieldbus LEDs]; see Chapter
.
-
PR 1721/41 is selected via
- [HW-Slots]- [Slot 2].
Display:
1
2
3
4
Constant
---
:
No function
Constant
grn
:
Module is online, da-
ta transmission is
possible.
Flashing 1 Hz
red
:
Input/output length
configuration error.
Constant
red
:
Fieldbus is offline,
data transmission is
not possible.
Flashing 2 Hz
red
:
User parameter error
Flashing 4 Hz
red
:
Error in ASIC
Legend
---
off
red
red
grn
green
4.6.5.3
Bus termination
The end nodes in a ProfiBus-DP network must be fitted with termination resistors, to
prevent reflections in the bus cable.
Bus termination switch
can only be accessed when the device
has been opened.
4 Device installation
Transmitter in field housing PR 5230
EN-81