4.5.8.2
Connecting load cells with a 6-wire cable
The load cell cable screens
are connected in the metal cable glands; see Chapter
The 6 load cell cores are color-coded:
Terminal
Port/color code
Description
M+
+ Meas./gn
+ Measuring voltage (load cell output)
M-
- Meas./gy
- Measuring voltage (load cell output)
+ Supply/rd
+ Supply voltage
V+
+ Sense/wh
+ Sense voltage
- Supply/bu
- Supply voltage
V-
- Sense/bk
- Sense voltage
Note:
The load cells are connected directly. Corner adjustment on the load cell junction board
is not provided.
4.5.9
Connecting an analog platform (series CAP…)
You can connect an analog platform to the device.
4 Device installation
Transmitter in field housing PR 5230
EN-71
Minebea Intec