Check the standard communica-
tion
↓
Communication of serial interface
is wrong.
→ Yes →
Move to check of serial interface
communication.
↓
No
↓
Standard RS-485 interface is
wrong.
→ No →
↓
Yes
↓
Inform Minebea Intec about the
contents of failure and situation in
details at site.
Is wiring correct?
→ No →
↓
Yes
↓
Connect surely with the external
equipment (refer to Chapter
Address No. is correctly set.
→ No →
↓
Yes
↓
Set the address No. correctly (refer
to Chapter
The communication protocol is
correctly set.
→ No →
↓
Yes
↓
Set it correctly according to the
speciications for host.
Setting of command/data is cor-
rect.
→ No →
↓
Yes
↓
Use correct command/data (refer
to Chapter
).
Host speciies the low control.
→ Yes →
↓
No
↓
Switch the low control at host side
from [YES] to [NONE].
Interval from the command receipt
to the transmission is 1 ms or more.
→ Yes →
↓
No
↓
Set the interval from the command
receipt to the transmission to 1 ms
or more.
Is the setting at the reply delay
time appropriate to the response
from host?
→ No →
↓
Yes
↓
The response from the host might
be slow.
Enlarge the response delay time of
this unit.
Inform Minebea Intec about the
contents of failure and situation in
details.
Weight Controller CSD-918
19 Troubleshooting
EN-194
Minebea Intec