16.4 Setting item of standard RS-485 communication
1
Operation mode
The operation mode is selectable from [COMMAND MODE] or [STREAM MODE].
[COMMAND MODE] is selected as default.
Note:
When stream mode is selected, the message of [DATA OUTPUT BY A CONSTANT
INTERVAL CONTINUES IN STREAM MODE.THEREFORE, CONTROL BY THE HOST
BECOMES IMPOSSI] is appeared. Tap [YES] to select.
2 Baud rate
Baud rate is selectable from [1 200 bps], [2 400 bps], [4 800 bps], [9 600 bps], [19
200 bps] or [38 400 bps].
[9 600 bps] is set as default.
3
Data bit length
The bit length of the output data is selectable from [7 bit] or [8 bit].
[8 bit] is selected as default.
4 Parity
The parity bit is selectable from [none], [Odd number] and [Even number].
[Odd number] is selected as default.
5 Stop bit
The stop bit is selectable from [1 bit] or [2 bit].
[1 bit] is selected as default.
6 Terminator
The terminator is selectable from [CR] or [CR+LF].
[CR+LF] is selected as default.
7
Addition of decimal point
The addition of decimal point is selectable from [None] or [Existing].
[None] is set as default.
8 Address
The address going to communicate is set. The address is input with the ten keys wit-
hin the range from 0 to 31.
[0] is set as default.
9 Data delay time of RS-485 interface
The data delay time when communicating with RS-485 interface is set.
After the host side transmission is inished, the return data of CSD-918 side can be
delayed in the case that sending terminal on the host side becomes the condition of
low impedance.
The time is input with the ten keys in the range from 0.00 to 9.99.
[0.01] is set as default.
10 Target output of stream mode
Data output at the time of stream mode is selectable from [LOAD DISPLAY],
[GROSS], [NET] or [LOAD DISP. WITH STATE].
[LOAD DISPLAY] is selected as default.
Weight Controller CSD-918
16 Standard RS-485 interface
EN-136
Minebea Intec