You can set the completion time of middle low in the range from [50 ms] to [10 000 ms],
and the inlection point C as the percentage between the completion time and AC of
middle low in the range from [0.00 %] to [100.00 %].
The default has set [2 000 ms] for the completion time of middle low and [50.00 %] for
the inlection point. In this case, the inlection point C is located after 1 000 ms from A.
AC=CB
O
pen
rat
e
[%]
Time [ms]
Inflection
point C
Middle Flow
Fine Flow
B
A
Full Flow
Inflection
point C
Middle Flow
Fine Flow
B
A
Full Flow
O
pen
rat
e
[%]
Time [ms]
AC>CB
O
pen
rat
e
[%]
Time [ms]
AC<CB
MiddleFlow
Inflection
point C
Fine Flow
B
A
Full Flow
Note:
In this control method, the gate operates continuously by always changing speed until
reaching the open rate set before at the set time.
11.4 Fine low (DRIBBLE FLOW)
The control method of Fine low is set.
The control method is [FIXED OPRN (WEIGHT)].
The open rate of the gate and the weight for [FIXED OPEN (WEIGHT)] are set.
1. In middle low screen, select [MEDIUM FLOW], and select [QUADRATIC EQUATION],
and then tap [OK].
2. Select [QUADRATIC EQUATION] in the middle low screen to inputting the each value.
3. Select [TIME] in quadratic control screen, and input the completion time of middle
low with ten keys, and then tap [OK].
4. Select [INFLECTION POINT], and input the percentage between A and C with ten keys,
and then tap [OK].
11 Weighing mode
Weight Controller CSD-918
Minebea Intec
EN-103