Page 59
Operation
CSP2008
Bit 0
Bit 1
Setup
active
< 500 ms
500 ms
Setup change
Switching inputs
Switching output
Fig. 48 Setup change by digital inputs
The digital inputs are transferred 500 milliseconds after concerning a stable level, see
Fig. 48
. In case of changing the level at the switching inputs during the setup change, it
does not result in a change of a setup selection. A new setup change is possible at the
earliest after 500 ms.
A setup change takes approximately 3 s without EtherCAT mode and approximately
10 s with EtherCAT mode.