Operation
CSP2008
5.3.11.5
Digital CSP2008
Synchronous output:
In this menu item, you define the pulse width of synchronization
output on the terminal block.
Statistics:
In this menu item, you define, which results of calculation program
Statis-
tics
are output to the two digital terminals
Out 1
and
Out 2, see Chap.
4.3.1
.
Possible results:
-
Block n - Statistics - Upper limit
|
Block n - Statistics - Lower limit
-
Block n - Statistics - Upper warning
|
Block n - Statistics - Lower warning
-
Block n - Statistics - IO, NIO
|
Block n - Statistics - Invalid meas value
Switching output active: HIGH
Switching output active: LOW
Hold time
Hold time
Switching
output
3.3 V
0 V
Time
Switching
output
3.3 V
0 V
Time
Fig. 44 Behavior of digital outputs DOUT 1 and DOUT 2
If the statistics module recognizes an invalid measurement value, no new statistic values
will be calculated.
i
If additional switching outputs are required, use EtherCAT terminals, see Chap.
5.3.11.6
.
Setup selection.
Alternatively to the statistics functions the used setup file or the state of
the controller can also be confirmed via the digital outputs.
Possible results:
-
Setup number bit 0
|
Setup number bit 1
-
Setup number bit 2
|
Setup number bit 3
-
Setup number bit 4
|
Setup active
If the controller changes a setup file, the
Setup active
signal is disabled.
Software digital outputs.
These outputs can also be confirmed via the digital outputs.
The software digital outputs are set in the
Settings > Digital CSP2008 outputs
menu or with a command, see Chap.
11.8
.
Possible results:
-
SW digital out 1
|
SW digital out 2
-
...
-
SW digital out 9
|
SW digital out 10
5.3.11.6
Digital EtherCAT
The digital EtherCAT terminals extend the choice of switching outputs.
Statistics:
In this menu item you define, which results of calculation program
Statis-
tics
are output to the digital terminals. The active behavior cannot be set at the digital
EtherCAT terminals.
Possible results:
-
Block n - Statistics - Upper limit
|
Block n - Statistics - Lower limit
-
Block n - Statistics - Upper warning
|
Block n - Statistics - Lower warning
-
Block n - Statistics - IO
|
Block n - Statistics - Invalid meas value
Switching output active: HIGH
Hold time
Switching
output
24 V
0 V
Time
Fig. 45 Behavior of digital outputs
If the statistics module recognizes an invalid measurement value, no new statistic values
will be calculated.