Page 13
Functional Principle, Technical Data
CSP2008
Calculation blocks
The routines for determining the thickness value and the master value with 10 mm are stored in the calculation
blocks.
Calculation
Block 1
Block 2
Program
Calculation
Mastering
Data sources
Data source 1
Sensor 1
Digital source 2
Sensor 2
Data source 1
Block 1 Calculation
Digital source 2
Zero
Calculation parameter
A = -1, B = -1, C = 0
Master value = 10 mm
Display
It is specified in the display settings which results should be displayed on the CSP2008.
Data output / Display
Thickness
Measured Value 1
Distance Sensor 1
Measured Value 2
Distance Sensor 2
Measured Value 3
Meass.Value 1 in mm
+010.000
Meass.Value 2
Meass.Value 3
+006.170 +007.804
Thickness value
Sensor 1 Sensor 2
Output
Block 2 Mastering Sensor 1
Sensor 2
Averaging
5
5
5
Analog output
Here you set the scaling and the output mode of the analog output.
Menu
Data output > Analog CSP2008
.
This sample program uses the analog
current output of 4 ... 20 mA for the master
value. With the defined output values
(minimum and maximum) this results in
an output current of 12 mA at the master’s
time.
Block 2 - Mastering
Output:
Hold last value
Error handling:
4 ... 20 mA
Output type:
Analog value 1
5.000000
Minimum output value in mm:
15.000000
Maximum output value in mm:
Ethernet
With the data output via Ethernet you define, which measured values will be presented in the Web interface or trans-
mitted via Ethernet to the PC.
Data output / Ethernet
Thickness
Measured Value 1
Distance Sensor 1
Measured Value 2
Distance Sensor 2
Measured Value 3
Output
Block 2 Mastering Sensor 1
Sensor 2
Saving
Saving of the parameter set, if you have made changes to these settings. Menu
Load/save settings > Load
CSP2008 setup from PC, see Chap.
5.3.12
.
Save the parameter set in a file, if required. Menu
Load/save settings > Load CSP2008 setup from PC,
see Chap.
5.3.12
. You will find the file <Setup(n).meo> in the desktop of your PC.
Start measurement
-
Put the master measuring object between the sensors.
-
Start mastering; therefore use the push button on the zero input.
i
You will find complete programs (setups) for the thickness measurement with laser-optical displacement sen-
sors and optical precision micrometers on the supplied CD or under
http://www.micro-epsilon.de/accessories/
Unicontroller.