Page 77
From EtherCAT to Ethernet
CSP2008
12.
From EtherCAT to Ethernet
Sensors or controller with active Ether-
CAT interface can be set in the op-
eration mode Ethernet via the
Ether
Mode
function.
Requirements:
-
The sensor or controller is con-
nected to the Sensor 1 / Sensor 2
connector.
-
RS422
must be selected in the
Sensor connection
.
-
The
baud rate
must be set cor-
rectly.
-
The EtherCAT Devices are set in the
operation mode
Sensor configu-
ration
.
opto
NCDT
ERR
P wer on
RUN
EtherCAT Ethernet
Laser off
In ange
Midrange
Error
F=1,5 . 50 kHz;
=670 nm
L se str h ung
i ht in den tr hl bli ken
L ser l sse 2
nach DIN N 60 25-1 2008-05
t=0 5. .542 µs
P =1,2 mW;
P
1 mW;
P
Sensor/Controller selection:
.
.
.
Charateristic
Sensor 1
ILD2300
Ether mode:
Ethernet
Measuring rate:
20 kHz
Sensor connection:
RS422
Baud rate:
691200
Submit the settings to CSP2008
EtherCAT bus:
Deactivated
Sensor configuration
Active
EtherCAT Devices
Submit
---
Bus coupler/ Sensor:
Procedure:
Change to the menu
Settings > Data recording > Sensor n
and select
Ethernet
for the
Ether Mode
.
Confirm your settings with
Submit
.
After the
Ether Mode
has been changed the sensor/controller will be reset. Afterwards
the sensor/controller will be automatically reconnected to the CSP2008.