Functional Description
26
b) from the screen
Example:
defines
the vertical scale
of the channel selected from the current
settings.
This yields a reading of the direct measurements of the quantity
analyzed and of its unit.
Coupling:
AC
AC
DC
DC
GND
GND
Coefficient:
Assignment of a multiplier coefficient to the sensitivity of
the selected channel using the stylus, on the digital keypad of the
"Coefficient" zone.
Validation by
.
The sensitivity indicated in the display of the parameters of the channel
will be modified as a function of this coefficient.
Unit of measurement:
Modification of the unit of the vertical scale of
the selected channel using the stylus in the table of usable characters
(not more than 3)
after the "measurement unit" zone has been chosen.
The unit of the vertical scale will be indicated in the display of the
parameters of the modified channel.
Bandwidth limit,
3 filters can be selected: 15MHz, 1.5MHz and 5kHz
BX limit is adjusted only from the adjustment menu of the channel, by
clicking it with the stylus
Limitation of the bandwidth of the channel and of its triggering circuit,
to moderate display noise and spurious triggerings.
The bandwidth of each channel can be limited to 5kHz, 1.5MHz, or
15MHz.
The limitation of the bandwidth of a channel is indicated in the
command zone by the parameter BW limit.
Selection of the colour:
-
red
-
green
-
magenta
-
blue