Applications
58
5.3 Calibrating the probes
Step
Action
1.
Connect the Probix adapter of an HX0030 probe
having a 1/10 ratio to the CH1 input.
2.
Connect the probe (with its ground) to the calibrator
output (Probe Adjust:
≈3
V,
≈
1kHz) on the side of the
instrument.
Connect the cold of the probe to the cold of the
calibration output of the probes.
4.
Check that the 1/10 coefficient of the probe has in
fact been taken into account.
Menu CH1
Click the right arrow, Measurement of probe,
select Coefficient: 10,
Validate by clicking "
»
Note: The sensitivity and the measurements take
the coefficient of the probe into account.
5.
Set the sensitivity of CH1.
Menu CH1, Sensitivity/coupling: 500mV/div
or using buttons A and B of the HX0030 probe
or using the
keys.
6.
Set the coupling of CH1.
Menu CH1, coupling: AC
or using the key.
7.
Set the sweep rate.
Time base menu: 500µs/div. or using the
keys.
8.
Set the triggering parameters
Trigg menu: Source: CH1, Coupling: AC, + Edge +
9.
Set the triggering mode.
Trigg Menu by the SGLE REFR. key
using the RUN HOLD key, start the acquisitions
("RUN" mode).
If necessary:
Modify the triggering level with the stylus by moving the T (Trigger) symbol on the screen. The triggering
level is indicated at bottom right on the screen.
Modify the vertical position of the curve by using the stylus to move the 1 symbol, to the left of the
screen.
The
key can be used to perform these adjustments automatically.