Remote Programing
101
* Command error:
(-199 to -100)
They indicate that a syntax error has been detected by the syntax analyzer and causes event
register bit 5, called CME, CoMmand Error to be set to 1.
-101 :
Invalid character
-103 :
Invalid separator
-104 :
Data type error
-108 :
Parameter not allowed
-109 :
Missing parameter
-111 :
Header separator error
-112 :
Program mnemonic too long
-113 :
Undefined header
-114 :
Header suffix out of range
-121 :
Invalid character in number
-128 :
Numeric data not allowed
-131 :
Invalid suffix
-138 :
Suffix not allowed
-141 :
Invalid character data
-148 :
Character data not allowed
-151 :
Invalid string data
-154 :
String data too long
-171 :
Invalid expression
* Execution errors:
(-299 to -200)
They indicate that an error has been detected at the moment of command execution and
causes event register bit 4, called EXE, Execution Error, to be set to 1.
-200 :
Execution error
-213 :
Init ignored
-221 :
Sandtings conflict
-222 :
Data out of range
-232 :
Invalid format
-256 :
File name not found
-257 :
File name error
* Specific instrument errors:
(-399 to -300)
They indicate that an abnormal error has been detected during execution of a task, and causes
event register bit 3, called DDE, Device Dependent Error to be set to 1.
-300 :
Device-specific error
-321 :
Out of memory
-350 :
Queue overflow
-360 :
Communication error
* Query errors:
(-499 to -400)
They indicate that an abnormal error has been detected during execution of a task, and cause
event register bit 2, called QYE, QuerY Error, to be set to 1.
-400 :
Query error
SYSTem:KLOCK (
Command/Query)
The
SYST:KLOCK <0|1|ON|OFF>
command locks the front face.
To the question
SYST:KLOCK?,
the instrument returns the lock status of the front face.
Summary of Contents for OX9302-BUS
Page 143: ...X04605B02 Ed 1 03 2018 ...