Applications
61
List of the different
values in
Auto measurements
Time measurements Level measurements
rise time DC voltage
fall time RMS voltage
positive pulse peak-to-peak voltage
negative pulse amplitude
duty cycle max. voltage
period min. voltage
frequency upper plateau
phase lower plateau
counting overshoot
integral
5.4.2. The cursors
There are three
categories
of cursors
(use the stylus
to move them).
Time cursors (T1 and T2), to measure certain time values and deduce a delta and its
frequency.
Amplitude cursors (V1 and V2), to measure amplitude values and deduce a delta.
A phase cursor to measure the phase of the signal according to the positioning of T1
and T2 and of a reference signal.
The phase cursor is inactive if you are in Auto measurement mode in all channels.
5.4.3. Zoom
For more accurate measurements with the cursors, press the key to use the Zoom function.
As default, the zoom is applied to the centre of the current acquisition of the
ScopiX
.
You can use the stylus to mark out a different zone.
The time base is corrected according to the zoom applied.
Zoomed screen
Press the key again to exit from the Zoom function.
Visually
complete
signal
Zoomed
signal