Functional Description
25
4.1.6
Adjustment of the amplitude of the "VERTICAL" signal
a) from the keypad
Selection of the channel
Activation of the channel
De-activation of the channel
Adjustment of the vertical
sensitivity
of the last channel selected:
Increases the vertical sensitivity
Decreases the vertical sensitivity
The sensitivity is indicated in the zone displaying the parameters of the channel.
It takes account of the parameters of the "Vertical scale" menu.
Adjustment of the
position
of the selected curve on the screen:
Move up
Move down
Selection by successive presses on the
input coupling,
"AC", "DC" or "GND", of the last
channel selected
Modification of the coupling
AC - DC - GND:
AC
blocks the DC component of the input signal,
attenuates signals below 10Hz.
DC
transmits the DC and AC components of the input signal.
GND
the instrument internally connects the input of the selected channel to a reference
level of 0V.
activates or deactivates the
horizontal
division by 4
of the display zone.
Activation of the "Full Trace" function is
indicated by:
the presence of a continuous horizontal
line between the display zones
horizontal division of the graticule by 2.
After activation of the function, the traces
can be moved vertically in their zones.