Functional Description
19
4. FUNCTIONAL DESCRIPTION OF OX 9304
4.1 SCOPE mode
4.1.1 Keys/ active keypad
4.1.2 Adjustment of the "Reference Memory" from the keypad
In oscilloscope mode, pressing this key freezes
the traces on the screen; the curve is displayed
in the channel colour, but darker, as reference to
be compared to a new acquisition.
The reference memories are accompanied by
their reference numbers.
Pressing again erases them: they are lost.
This memory is not saved and will be lost when
you exit the Oscilloscope mode.
4.1.3 Adjusting the AUTOSET from the keypad
"Magic Wand" key
Automatic optimum adjustment of the AUTOSET of the channels to which a signal is applied.
The adjustments concerned are:
the coupling
the vertical sensitivity
the time base
the slope
the positions
the triggering.
The signal having the lowest frequency is used as triggering source.
If no trace is detected on the inputs, autoset is aborted.
A simultaneous press on
+
assigns the corresponding channel as triggering source.
"Home"
Automatic
measurements
Reference
of the
measurements
Cursors
ON/OFF key
Horizontal
time base
Zoom
Vertical
adjustments
Triggering
Brightness
Full screen
Screen shot
Reference
memory
Autoset