General Characteristics
81
9.2. Mechanical characteristics
9.2.1. Hard enclosure covered with elastomer
Comprising
:
- a lower housing,
- a central belt holding all terminations,
- an upper housing,
- a battery compartment cover.
Dimensions:
292.5x210.6x66.2mm
Weight:
approximately 2.4 kg with the battery
Carrying strap:
snaps onto the top of the instrument
9.2.2. Mechanical conditions
Tightness
Tight to drops of water falling vertically and penetration of objects
≥
1mm: IP 54 (instrument not in
operation)
Instrument alone, without accessories or mains power supply, upright, tilted 40° on its prop or flat with LCD
up.
Remarks:
1. Do use not the instrument in a atmosphere laden with carbon dust, metallic dust,
or other conducting dust.
2. Wipe the instrument, in particular the measurement terminals, before using
again.
Shocks and impacts
Per the test standards of IEC 62262: IK03 (LCD screen) and IK06 (any other part of the instrument)
3 impacts with an energy of 1 Joule (IK06) or 0.35 Joule (IK03), applied to each component part of the
instrument, without deterioration that might create a risk for the safety of the user.
Free
fall
, without packaging.
Instrument alone, without accessories, on 3 sides.
Per the test standards of IEC 61010-1-2010.
Summary of Contents for OX9302-BUS
Page 143: ...X04605B02 Ed 1 03 2018 ...