Applications
65
Remarks
The channels of the measurement ranges are automatic. To define the measurement range in
manual mode, press the key opposite.
A long press on the key of the channel is used to return to automatic mode. In addition:
in automatic mode, the measurement range on the screen is highlighted in the colour of the
channel
in manual mode, it is not.
The coupling of the channels can be modified using the key opposite:
DC
AC
AC <5kHz
AC <625kHz
AC+DC
AC+DC <5kHz
AC+DC <625kHz
6.3 Power measurement
To measure power, you must have the right
Probix
accessories:
current measurements are made using
HX0034
,
HX0072,
and
HX0073
clamps
voltage measurements are made using the
HX0033
banana adapter and leads.
A power measurement is made in the Multimeter mode, by clicking the
icon. Then, select the type of set-up you want
to measure:
Single-phase power
Display of the result of calculation of the active power, measured
using CH1 for the voltage measurement and CH4 for the current
measurement.
Three-phase power
on balanced network
without neutral
The value displayed is the active three-phase power calculated from
the wiring proposed at the time of selection.
Three-phase power
on balanced network
with neutral
The value displayed is equal to 3 times the active power measured on
one phase.
Three-phase power,
3 wires
Display of the result of calculation of the active three-phase power
measured by the two-wattmeter method on a installation without
neutral.
When the values are read in this mode, the following screen is displayed:
Example
: Single-phase power
Channel 1 indicates the
voltage
measured
directly with its min and max values
Channel 4 indicates the
current
measured
directly with its min and max values.
The various
power values
calculated from
channels 1 and 4 are displayed, along with
their
power factor
.
The type of wiring is indicated next to the
values.
Summary of Contents for OX9302-BUS
Page 143: ...X04605B02 Ed 1 03 2018 ...