Remote Programing
95
MEASure:DMM?
(Query)
To the question
MEAS:DMM? <INT1|2|3|4>
the instrument returns the value of the main
measurement for the selected channel.
INT1 to INT4 index are associated with channels 1 to 4. Use the index to find INT5 power
measurement.
Before using the command
MEAS: DMM? INT5
, the instrument must be configured to
measure the power measurement (see [SENSe]: Function).
Response format :
<mesure> <NL>
value format <nrf>
MEASure:FALL:OVERshoot?
(Query)
To the question
MEAS:FALL:OVER? <INT{1|2|3|4}>
the instrument returns the negative
overshoot of the selected signal.
Response format:
<measured value><NL>
value in format <NR2> expressed in percent.
MEASure:FALL:TIME?
or
MEASure:FTIME?
(Query)
To the question
MEAS:FALL:TIME? <INT{1|2|3|4}>
the instrument returns the fall time of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
MEASure:FREQuency?
(Query)
To the question
MEAS:FREQ? <INT{1|2|3|4}>
the instrument returns the frequency of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in hertz.
MEASure:HIGH?
(Query)
To the question
MEAS:HIGH? <INT{1|2|3|4}>
the instrument returns the value of the high level
level of the selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
MEASure:LOW?
(Query)
To the question
MEAS:LOW? <INT{1|2|3|4}>
the instrument returns the low level value of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
MEASure:MANual:PHASe?
(Query)
To the question
MEAS: MAN:PHAS?,
the instrument returns the phase of
ϕ
-cursor in relation
to cursors 1 and 2. The difference between the cursor 1 and 2 represents 360°. The cursor 1
equal to 0° and the cursor 2, 360°.
Response format:
<measured value><NL>
value in format <NR2> expressed in degree.
MEASure:MAXimum?
(Query)
To the question
MEAS:MAX? <INT{1|2|3|4}>
the instrument returns the maximum value of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
MEASure:MINimum?
(Query)
To the question
MEAS:MIN? <INT{1|2|3|4}>
the instrument returns the value minimum of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
Summary of Contents for OX9302-BUS
Page 143: ...X04605B02 Ed 1 03 2018 ...