How to store the XCi when not in use
Although the vast majority of installed HydroMACE XCi devices are left "in the field" during periods of non-
use, some users choose to remove them to "overwinter".
If removing the device, there are some simple steps to make sure everything will work upon the next
installation:
Store the XCi
1. Always "
Stop
" the unit with
2.
ALWAYS
leave the main battery connected.. The XCi draws very low current (µA) when "stopped",
and the main battery will last at least 6 months.
NOTE: If the XCi is not going to be used for more than 6 months, the XCi
device should be put on a 16-30VDC charger periodically to keep the battery
charged.
Store the sensor
1. If removing the sensor, ensure that it is stored somewhere where the connector end of the sensor
remains dry
AT ALL TIMES
.
2. If you have an area/velocity sensor, "
Disable
" the reference filter by firmly screwing the cap closed.
3.
WARNING: If you intend to leave the sensor in the field without being
connected to the XCi device, you MUST make the connector end of the sensor
waterproof.
MACE XCi User Manual - 262 -