168
47DAC OPERATING MANUAL EM-9518-B Rev.4
• TABLE 6: NORMAL MODE
TIMING signal
When the TIMING signal is ON, ‘TG’ indicator turns on,
however the signal is invalid. Measures continuously.
HOLD
RESET
*
2
*
3
Hold
*
1 Scaling error
*
2 Measuring
*
3 Not measuring
Display
reading
Scaling error range
Hold
*
2
*
1
*
2
HOLD signal
While the HOLD signal is ON, the display reading is held
at the measured value at the falling edge of the signal,
and the TIMING, S-TMR and ZERO signals are invalid.
Scaling error
During the scaling error, data measuring continues.
However the display shows the scaling error. If the
HOLD signal turns ON during the scaling error and is
maintained after the scaling error is canceled, the display
reading changes to the measured value at the falling
edge of the signal.
RESET signal
Resets the measured data with the RESET signal ON
(no measurement).
DC output
Proportional to the display reading. For no measured
data, the value is 0%.
• TABLE 7: SAMPLING HOLD MODE
TIMING signal
Data is measured at the falling edge of the TIMING
signal.
TIMING
RESET
*
1 Scaling error
*
1
Display
reading
Scaling error range
Hold
Hold
Hold
Not measuring
Not measuring
TIMING within RESET: Invalid
Not measured if TIMING rises during
a scaling error
HOLD signal
While the HOLD signal is ON, the display reading is held
at the measured value at the falling edge of the signal,
and the TIMING, S-TMR and ZERO signals are invalid.
Scaling error
Stops data measuring.
RESET signal
Resets the measured data with the RESET signal ON
(no measurement). The TIMING signal is invalid.
DC output
Proportional to the display reading. For no measured
data, the value is 0%.
• TABLE 8: PEAK HOLD MODE
TIMING signal
While the TIMING signal is ON, the MAX value within the
period is measured and stored, and is established at the
rising edge of the signal.
*
1 Scaling error
*
2 Measuring
TIMING
RESET
*
2
*
2
*
1
Display
reading
Scaling error range
Hold
Hold
Not measuring
Not measuring
TIMING within RESET: Invalid
Not measured if a scaling error occurs
while TIMING is ON
HOLD signal
While the HOLD signal is ON, the display reading is
held at the last measured value at the falling edge of the
signal, and the TIMING, S-TMR and ZERO signals are
invalid.
Scaling error
Stops data measuring.
RESET signal
Resets the measured data with the RESET signal ON
(no measurement). The TIMING signal is invalid.
DC output
Proportional to the display reading. For no measured
data, the value is 0%.
• TABLE 9: VALLEY (BOTTOM) HOLD MODE
TIMING signal
While the TIMING signal is ON, the MIN value within the
period is measured and stored, and is established at the
rising edge of the signal.
*
1 Scaling error
*
2 Measuring
TIMING
RESET
*
2
*
2
*
1
Display
reading
Scaling error range
Hold
Hold
Not measuring
Not measuring
TIMING within RESET: Invalid
Not measured if a scaling error occurs
while TIMING is ON
HOLD signal
While the HOLD signal is ON, the display reading is
held at the last measured value at the falling edge of the
signal, and the TIMING, S-TMR and ZERO signals are
invalid.
Scaling error
Stops data measuring.
RESET signal
Resets the measured data with the RESET signal ON
(no measurement). The TIMING signal is invalid.
DC output
Proportional to the display reading. For no measured
data, the value is 0%.