7
SERVICE
64
OPTIWAVE 1010
www.krohne.com
10/2016 - 4003537403 - MA OPTIWAVE 1010 R03 en
7.2.2 How to replace the signal converter
Equipment needed:
•
5 mm Allen wrench (not supplied)
•
OPTIWAVE 1010 level transmitter attached to a BM 26 Advanced bypass level indicator
•
New housing and electronics block. For more data, refer to
Spare parts on page 86. Speak to
your supplier to get a new housing and electronics block.
•
Ferrite choke. For more data, refer to
Electrical installation: 2-wire, loop-powered on page
26.
•
Handbooks for all devices
CAUTION!
Make sure that you record device settings. For more data about how to save device settings,
refer to Import / Export on page 50.
Figure 7-1: How to remove the signal converter