OPERATION
6
59
OPTIWAVE 1010
www.krohne.com
10/2016 - 4003537403 - MA OPTIWAVE 1010 R03 en
6.7.9 DTM settings
This is the menu for the device status display at the top of the DTM window. Use this menu to
start and stop status checks of the device and change the interval between checks.
Function description
Figure 6-20: DTM settings menu
1
Status display (device status)
2
Last recorded measurement
3
Menu item: status display enabled (on) / disabled (off)
4
Menu item: update interval
INFORMATION!
Click on the button below the device status symbol to change from distance to level
measurement.
Function
Function description
Selection list or range
of values
Default
Status display
This menu item starts and stops device status
checks. If the device operates correctly, the status
display at the top right side of the window will show a
green screen with a tick symbol. If the device is not
connected to the computer or this menu item is set to
"disabled", the status display will show a gray screen.
Enabled, Disabled
Enabled
Update interval
This menu item changes the interval between device
status checks.
s = seconds
min-max:
15
…
3600 s
30 s