Keysight N5247B Service Guide 3-63
Tests and Adjustments
Adjustments
a.
In “Select Adjustments,” leave all four adjustments selected by
default.
b.
In “Power Meter”
—
Set the GPIB address to match the address of the power
meter.
—
If using a single channel power meter, use the default setting
Ch. A. If using a dual channel power meter and the sensor is
attached to the B channel, select Ch. B.
—
If the sensor has already been zeroed and calibrated, select
Skip Zero and Cal before first use.
Figure 3-21
PNA Noise Adjustment Dialog Box
3.
Click Begin.
4.
As shown in
, connect an RF cable between the network
analyzer test ports 1 and 2.