Keysight N5247B Service Guide 3-5
Tests and Adjustments
Before You Begin
Allow the Analyzer to Warm Up
Review the Principles of Connector Care
Proper connector care and connection techniques are critical for accurate and
repeatable measurements. Refer to
for tips on connector care.
Prior to making connections to your analyzer, carefully review the information
about inspecting, cleaning, and gaging connectors. Refer to the calibration kit
documentation for detailed connector care information.
For course numbers about additional connector care instruction, contact
Keysight Technologies. Refer to
“Contacting Keysight” on page 2-7
To achieve the maximum system stability, allow the analyzer to warm up
for at least 90 minutes.
Cleaning connectors with alcohol shall only be done with the
instrument power cord removed and in a well-ventilated area. Allow
all residual alcohol moisture to evaporate, and the fumes to
dissipate prior to energizing the instrument.
Table 3-1
Connector Care Quick Reference Guide
Handling and Storage
Do
— Keep connectors clean
Do Not
— Touch mating-plane surfaces
— Extend sleeve or connector nut
— Set connectors contact-end down
— Use plastic end-caps during storage
— Store connectors or adapters loose
Visual Inspection
Do
— Inspect all connectors carefully
Do Not
— Use a damaged connector - ever
— Look for metal particles, scratches, and
dents
Connector Cleaning
Do
— Try compressed air first
Do Not
— Use any abrasives
— Use isopropyl alcohol
a
• Get liquid into plastic support beads
— Clean connector threads
Gaging Connectors
Do
— Clean and zero the gage before use
Do Not
— Use an out-of-specification
connector
— Use the correct gage type
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...