Tests and Adjustments
Adjustments
3-
4.
Follow the instructions and prompts as they are displayed.
IF Gain Adjustment
The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then
click IF Gain Adjustment.
2.
Follow the instructions and prompts as they are displayed.
Receiver Characterization
This characterizes the receivers in your analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then
click Receiver Characterization.
2.
Follow the instructions and prompts as they are displayed.
Receiver Adjustment
The receiver calibration is used to adjust the network analyzer receivers for a
flat response across its full frequency range:
1.
A power meter/sensor is connected to Port 1, as shown in
, to
establish a reference for flatness.
2.
A cable is inserted between the power sensor and the test port, as shown
in
, to establish a reference for the cable.
3.
The same cable is connected between test port 1 and test port 2, as
, and a signal from Port 1 is used to adjust the “B”
receiver at Port 2.
The adjustment is repeated using a signal from Port 2 to adjust the “A”
receiver at Port 1.
Data obtained during this adjustment are stored in the mxcalfile_pxx files on
the solid state drive. The data are used in subsequent measurements.
If the solid state drive is replaced, these mxcalfile_pxx files will be lost.
Therefore, they should be backed up (saved on a floppy disk) so that they can
be restored. If using multiple disk drives (e.g. classified and general usage),
then these files must be replaced on each individual disk drive.
These files can be recreated by performing another receiver calibration
adjustment.
System
Service
Adjustment Routines
System
Service
Adjustment Routines
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...