5-64 Keysight N5247B Service Guide
Theory of Operation
Power Supply Group Operation
5-
Power Supply Group Operation
The A20 power supply assembly is a switching power supply operating at 103
kHz switching frequency. The input power ranges for the power supply are 90
to 132 Vac or 195 to 250 Vac. The power supply automatically senses the input
voltage and switches between these two ranges.
The dc output voltages of the A20 power supply assembly are:
— +15 V analog
— +9 V analog
— +3.3 V analog
— +5.2 V analog
— +15 V standby (always on)
— +32 V analog
—
−
15 V analog
—
−
5.2 V analog
—
−
7 V analog
— +5.1 V standby
— +12 V digital
— +3.35 V digital
— +5.1 V digital
The +15 V standby supply remains on continuously whenever the power supply
is plugged in. This supply is used to provide power to front panel LEDs and
CPU components when the analyzer is turned off.
Supply voltages which oscillate between the two normal input
ranges of the autoranging line voltage input will damage the power
supply. In rare cases, this damage has become a user safety
concern. If unstable power levels are expected, the analyzer input
power must be buffered by a line conditioner.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...