Keysight N5247B Service Guide 3-57
Tests and Adjustments
Adjustments
Equipment Used for the Source Adjustment
Procedure
1.
Connect the equipment as shown in
. Connect a GPIB cable
between the network analyzer and the power meter.
Figure 3-17
Equipment Setup for the Source Adjustment
2.
.Press UTILITY
, then
, then
, then
click Source Adjustment.
3.
Ensure the GPIB settings are correct.
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, Type-N
8482A
None
Power sensor, 2.4 mm
8487A
None
Power sensor, V-Band
a
a. This sensor and adapter must be calibrated together by the Keysight factory. The data supplied is
only valid as long as the sensor and adapter remain connected.
V8486A
None
Adapter, WR-15 to 1.85 mm (m)
V281B
None
Adapter, 1.85 mm (f) to 1.85 mm (f)
85058-60114
None
Adapter, 2.4 mm (f) to 2.4 mm (f)
11900B
85056-60007
Adapter, Type-N (f) to 2.4 mm (f)
b
b. The 1.85 mm connector has the same ruggedness and is compatible with the 2.4 mm connectors.
1.85 mm and 2.4 mm adapters are interchangeable.
11903B
None
System
Service
Adjustment Routines
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...