Keysight N5247B Service Guide 7-49
Repair and Replacement Procedures
Removing and Replacing the A37 Reference Mixer Switch
When reinstalling the A37 reference mixer switch into the analyzer, loosely
install the two mounting screws (item
). Connect the four semirigid
cables (item
) and torque the connectors to 10 in-lbs, then tighten the
two mounting screws (item
).
2.
Perform the post-repair adjustments, verifications, and performance tests
that pertain to this removal procedure. Refer to
.