4-46 Keysight N5247B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
—
W30 on the A63 doubler board
2.
Disconnect cable W27 or W28 or W29 or W30, depending on which
doubler board is to be checked.
3.
Connect the spectrum analyzer to the open connector on the doubler
board to be checked.
4.
Set the PNA for an 800 MHz CW output frequency and observe the
spectrum analyzer display. An 800 MHz signal should be present.
5.
Refer to
. If the PNA problem you are
troubleshooting is frequency banded rather than broadband related, set
the PNA output frequency to the center of the problem band. The
spectrum analyzer should display a signal above the PNA setting. For
analyzers with DSP version 4.0, the signal is 7.61 MHz above the network
analyzer setting. For analyzers with DSP version 5.0, the signal is 7.44
MHz above the network analyzer setting.
6.
If the signal is not present but the signals described in
and A17 Source Synthesizer Outputs” on page 4-43
are present, replace
the appropriate doubler board. Refer to
A60–A63 70 GHz Doublers” on page 7-62
.
7.
If the signal is present, reconnect all cables, and then continue with
“Checking the Signal through the Signal Separation Path” on page 4-49
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...