5-48 Keysight N5247B Service Guide
Theory of Operation
Signal Separation Group Operation
5-
The coupled arm of the couplers carries the signal reflected from or
transmitted through the DUT, to the receiver for measurement (through front
panel jumpers), as inputs A, B, C, and D for 4-port models or inputs A and B for
2-port models. The coupling coefficient of the directional couplers is nominally
15 dB for all frequencies above 500 MHz. The coupling coefficient increases for
frequencies below 500 MHz.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...