Keysight N5247B Service Guide 5-3
Theory of Operation
Network Analyzer System Operation
Network Analyzer System Operation
The PNA network analyzer generates two (2-port models) or four (4-port
models) phase-locked incident signals and an LO signal from the internal
synthesized source. By means of signal separation, the incident signals are
divided into reference signals and test signals.
The reference signals are applied to the receiver group, while the test signals
are applied to the device under test (DUT) and then to the receiver group. The
LO signal is applied directly to the receiver group where it is mixed with the test
and reference signals to produce IF signals for each of the eight receivers (A–D,
R1–R4) for 4-port models or four receivers (A, B, R1, R2) for 2-port models.
These IF signals are downconverted and then sampled and digitally processed.
If configured with the optional 2nd source, the 2-port models provide two
auxiliary source outputs, SRC 2 OUT 1 and SRC 2 OUT 2.
is a simplified block diagram of the 4-port network analyzer system
is a simplified block diagram of the 2-port network analyzer
system.
Figure 5-1
4-Port System Simplified Block Diagram
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...