3-6 Keysight N5247B Service Guide
Tests and Adjustments
Before You Begin
3-
— Use correct end of calibration block
— Gage all connectors before first use
Making Connections
Do
— Align connectors carefully
Do Not
— Apply bending force to connection
— Make preliminary connection contact
lightly
— Over tighten preliminary connection
— Turn only the connector nut
— Twist or screw any connection
— Use a torque wrench for final
connection
— Tighten past torque wrench “break”
point
a. Cleaning connectors with alcohol shall only be done with the instrument’s power cord removed, and in a well-ventilated
area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument.
Table 3-1
Connector Care Quick Reference Guide