Keysight N5247B Service Guide 3-9
Tests and Adjustments
About System Verification and Performance Tests
Certificate of Calibration
Keysight Technologies will issue a certificate of calibration upon successful
completion of system verification or completion of the performance tests. The
certificate of calibration will apply to the “system” (analyzer, calibration kit, test
cables, and any necessary adapters) if the system verification procedure is
used to confirm the system specifications. If the performance tests are used to
confirm instrument specifications, the certificate of calibration will apply to the
PNA as an independent instrument. The equipment and measurement
standards used for the tests must be certified and must be traceable to
recognized standards.
If you have a measurement application that does not use all of the
measurement capabilities of the analyzer, you may ask your local Keysight
Technologies service office to verify only a subset of the specifications.
However, this “limited calibration” creates the possibility of making
inaccurate measurements if you then use the analyzer in an application
requiring additional capabilities.
Summary of Contents for N5247B
Page 10: ...10 Keysight N5247B Service Guide Contents ...
Page 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Page 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Page 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Page 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Page 565: ...Installation Note Xxxxx xxxxx 3 ...